Development of a semi-empirical method to determine the efficiency of a gamma radiation detector for point sources

Main Article Content

Pablo Ortiz-Ramírez
Franco López-Usquiano

Abstract

 

In the present work, we introduce the theoretical development and a new approach to validate a novel method to calibrate the efficiency of a gamma radiation detector for point sources, which we have named the “Efficiency Extrapolation Method”.

The method consists in the determination of the detector efficiency using a set of monoenergetic gamma sources, which we will refer to as reference efficiencies. From these values, we will extrapolate the detector efficiency to the complete energy range using the first principle physics of gamma radiation detection theory. Therefore the proposed method corresponds to a semi-empirical one.

The determination of reference efficiencies must be done experimentally, but in this work, simulations were performed using FLUKA code. The reference energies will be: 59,54 keV, 661,66 keV and 1 274,54 keV associated to the isotopes 241Am, 137Cs and 22Na respectively. The second part of the method, the extrapolation from the reference energies to the gamma range, will be done over the energies emitted by 152Eu and 133Ba.

In general, the results are very good. While the obtained results for energies 53,16 keV and greater than 344,3 keV show an excellent agreement, the results obtained for energies in the middle range are only good.

 

Article Details

How to Cite
Ortiz-Ramírez, P., & López-Usquiano, F. (2019). Development of a semi-empirical method to determine the efficiency of a gamma radiation detector for point sources. Nucleus, (65), 47-50. Retrieved from http://nucleus.cubaenergia.cu/index.php/nucleus/article/view/679
Section
Ciencias Nucleares

References

[1] BÖHLEN TT, CERUTTI F, CHIN MPW, FASSO A, et. al. The FLUKA code: developments and challenges for high energy and medical applications. Nuclear Data Sheets. 2014; (120): 211-214.
[2] BERGER MJ, HUBBELL JH, SELTZER SM, CHANG J, et. al. XCOM: photon cross section database (version 1.5). [Online] Available: http://physics.nist.gov/xcom.
[3] ORTIZ RAMÍREZ P, LÓPEZ USQUIANO F. Development of a semi-empirical method to determine the efficiency of a gamma radiation detector. Proceedings del XX Simposio de la Sociedad Chilena de Física. Santiago, Chile. November 30- December 1-2, 2016.
[4] MOENS L, DE DONDER J, XI-LEI L, et. al. Calculation of the absolute peak efficiency of gamma-ray detectors for different counting geometries. Nucl Instrum & Meth in Phys Res. 1981; 187(2-3): 451-472.