Patient-specific intensity modulated radiotherapy quality assurance using an electronic portal imaging device
Main Article Content
Abstract
modern clinical electron linear accelerator was commissioned at the Department of Radiotherapy in the Oncology and Radiobiology Institute. The use of electronic portal image device (EPID) for patient-specific intensity modulated radiotherapy (IMRT) quality assurance is assessed and validated. An IMRT step & shoot approach, using multileaf collimator was used. The EPID dose-response curve was determined for a 6MV photon beam. The EPID effective water scattering depth () was assessed through measurements of its response versus field size variation and compared with the output factor scatter at different depths, calculated with a radiotherapy treatment planning system (TPS).
Dose profiles and dose maps were calculated at with the TPS, using a collapsed beam setup. The dose maps and profiles were compared with the processed EPID images for simple cases (i.e. open and wedged fields) and a more complex case (a head and neck IMRT clinical case).
Comparison was performed using the Gamma function. Results showed Gamma passing rates higher than 90% for all compared points, using comparison criteria of dose difference 3% and 4mm of distance to agreement.
Dose profiles and dose maps were calculated at with the TPS, using a collapsed beam setup. The dose maps and profiles were compared with the processed EPID images for simple cases (i.e. open and wedged fields) and a more complex case (a head and neck IMRT clinical case).
Comparison was performed using the Gamma function. Results showed Gamma passing rates higher than 90% for all compared points, using comparison criteria of dose difference 3% and 4mm of distance to agreement.
Article Details
How to Cite
Larrinaga Cortina, E. F., & Alonso Fernández, D. N. (1). Patient-specific intensity modulated radiotherapy quality assurance using an electronic portal imaging device. Nucleus, (59). Retrieved from http://nucleus.cubaenergia.cu/index.php/nucleus/article/view/620
Section
Ciencias Nucleares
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
Aquellos autores/as que tengan publicaciones con esta revista, aceptan los términos siguientes:
- Los autores/as conservarán sus derechos de autor y garantizarán a la revista el derecho de primera publicación de su obra, el cuál estará simultáneamente sujeto a la Licencia Creative Commons Attribution-NonCommercial 4.0 International (CC BY-NC 4.0) que permite a terceros compartir la obra siempre que se indique su autor y su primera publicación esta revista. Bajo esta licencia el autor será libre de:
- Compartir — copiar y redistribuir el material en cualquier medio o formato
- Adaptar — remezclar, transformar y crear a partir del material
- El licenciador no puede revocar estas libertades mientras cumpla con los términos de la licencia
Bajo las siguientes condiciones:
- Reconocimiento — Debe reconocer adecuadamente la autoría, proporcionar un enlace a la licencia e indicar si se han realizado cambios. Puede hacerlo de cualquier manera razonable, pero no de una manera que sugiera que tiene el apoyo del licenciador o lo recibe por el uso que hace.
- NoComercial — No puede utilizar el material para una finalidad comercial.
- No hay restricciones adicionales — No puede aplicar términos legales o medidas tecnológicas que legalmente restrinjan realizar aquello que la licencia permite.
- Los autores/as podrán adoptar otros acuerdos de licencia no exclusiva de distribución de la versión de la obra publicada (p. ej.: depositarla en un archivo telemático institucional o publicarla en un volumen monográfico) siempre que se indique la publicación inicial en esta revista.
- Se permite y recomienda a los autores/as difundir su obra a través de Internet (p. ej.: en archivos telemáticos institucionales o en su página web) antes y durante el proceso de envío, lo cual puede producir intercambios interesantes y aumentar las citas de la obra publicada. (Véase El efecto del acceso abierto).
La Revista Nucleus solo aceptará contribuciones que no hayan sido previamente publicados y/o procesados, por otra publicación. Cualquier violación ese sentido será considerada una falta grave por parte del autor principal lo cual será objeto valoración por parte del Consejo Editorial, el cual dictaminará al respecto.
References
[1] EUROPEAN SOCIETY FOR RADIOTHERAPY AND ONCOLOGY. Guidelines for the verification of IMRT. Booklet No.9. Belgium: ESTRO, 2008.
[2] International Commission on Radiation Units and Measurements. Prescribing, Recording, and Reporting Photon-Beam Intensity-Modulated Radiation Therapy (IMRT). ICRU Report 83. United Kingdom: Oxford Press, 2010.
[3] LEE C, MENK F, CADMAN P, et. al. A simple approach to using an amorphous silicon EPID to verify IMRT planar dose maps. Med Phys. 2009; 36(3): 984-992.
[4] LIEBICH J, LICHER J, SCHERF C, et. al. Simple proposal for dosimetry with an Elekta iViewGTTM Electronic Portal Imaging Device (EPID) Using Commercial Software Modules. Strahlenther Onkol 2011; 187(5): 316-321.
[5] VAN ELMPT W, MCDERMOTT L, NIJSTEN S, et. al. A literature review of electronic portal imaging for radiotherapy dosimetry. Radiother Oncol. 2008; 88(3): 289-309.
[6] SILVESTRE PATALLO I, GARCÍA YIP F. Uso del sistema imagen portal electrónica iViewGT para control de calidad del haz de radiación de un acelerador lineal de uso clínico. Revista Cubana de Física 2010; 27(2B): 251-256.
[7] HERMAN MG, BALTER J, JAFFRAY D, MCGEE K, et. al. Clinical use of electronic portal imaging: report of AAPM Radiation Therapy Committee Task Group 58. Med Phys. 2001; 28(5): 712-737.
[8] National Institute of Health. USA. Image Processing and Analysis in Java (ImageJ) [website]. Available in: https://imagej.nih.gov/ij.
[9] LOWE D, HARMS W, MUTIC S, PURDY J. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998; 25(5): 656-661.
[10] Elekta Oncology Systems. iViewGTTM Corrective Maintenance Manual. 2008.
[11] Organismo Internacional Energía Atómica. Aspectos físicos de la garantía de calidad en radioterapia: Protocolo de control de calidad. OIEA-TECDOC 1151. Viena: OIEA, 2000.
[2] International Commission on Radiation Units and Measurements. Prescribing, Recording, and Reporting Photon-Beam Intensity-Modulated Radiation Therapy (IMRT). ICRU Report 83. United Kingdom: Oxford Press, 2010.
[3] LEE C, MENK F, CADMAN P, et. al. A simple approach to using an amorphous silicon EPID to verify IMRT planar dose maps. Med Phys. 2009; 36(3): 984-992.
[4] LIEBICH J, LICHER J, SCHERF C, et. al. Simple proposal for dosimetry with an Elekta iViewGTTM Electronic Portal Imaging Device (EPID) Using Commercial Software Modules. Strahlenther Onkol 2011; 187(5): 316-321.
[5] VAN ELMPT W, MCDERMOTT L, NIJSTEN S, et. al. A literature review of electronic portal imaging for radiotherapy dosimetry. Radiother Oncol. 2008; 88(3): 289-309.
[6] SILVESTRE PATALLO I, GARCÍA YIP F. Uso del sistema imagen portal electrónica iViewGT para control de calidad del haz de radiación de un acelerador lineal de uso clínico. Revista Cubana de Física 2010; 27(2B): 251-256.
[7] HERMAN MG, BALTER J, JAFFRAY D, MCGEE K, et. al. Clinical use of electronic portal imaging: report of AAPM Radiation Therapy Committee Task Group 58. Med Phys. 2001; 28(5): 712-737.
[8] National Institute of Health. USA. Image Processing and Analysis in Java (ImageJ) [website]. Available in: https://imagej.nih.gov/ij.
[9] LOWE D, HARMS W, MUTIC S, PURDY J. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998; 25(5): 656-661.
[10] Elekta Oncology Systems. iViewGTTM Corrective Maintenance Manual. 2008.
[11] Organismo Internacional Energía Atómica. Aspectos físicos de la garantía de calidad en radioterapia: Protocolo de control de calidad. OIEA-TECDOC 1151. Viena: OIEA, 2000.